SKY1016-IV Aviation Fuel Automatic Freezing Point Tester
Product Introduction
SKY1016-IV Automatic Freezing Point Tester for Aviation Fuel is suitable for testing the freezing point of aviation fuel to help determine the low temperature flow properties of aviation fuel. The instrument uses automatic phase conversion method for rapid testing, accurate, efficient, fully automatic "one-button" test, and is ideal for laboratory aviation fuel freezing point determination.
Applicable standards: ASTM D5972, SH/T 0770, IP 435

The instrument adopts iSKvator technology, one-button start, and the test process of purging, cooling, constant temperature, heating, result judgment, etc. is fully automatic.


It effectively liberates manpower and reduces human errors. The automatic phase conversion method is used for freezing point detection, and the test results are consistent with the results measured by GB/T 2430 and ASTM D2386 methods.


The instrument is ergonomically designed and uses a 12" touch screen computer with a beautiful interface, simple operation, rich information, and real-time display of temperature and optical signal curves.


The test process is intuitive and clear.


The test only requires 0.14mL~0.16mL of micro-samples, and the test time is about 15 minutes. It is easy to clean, fast and efficient.


Special Peltier/compressor hybrid refrigeration technology and intelligent temperature control ensure high efficiency and low depth refrigeration.


Automatically collect test data: oil name and number, experimental time, purge timing, test results, etc.


The refrigeration test program can evaluate the refrigeration effect of the instrument at any time, effectively ensuring the test accuracy of the instrument.


The structural design is compact and reasonable, and the test consumables and wearing parts are easy to replace. It can store Less than 5,000 test result records, and provides data query, printing, deletion and other management functions


Using professional optical sensors with high sensitivity, high stability and high anti-interference ability, it can accurately and quickly detect the freezing point of oil samples

PT100 platinum resistance temperature probe to ensure high-precision temperature control of the instrument


High-performance special drying components effectively filter air moisture, no longer need to use laboratory compressed gas, improve the safety and flexibility of instrument use


Temperature sensors, solenoid valves, Peltier and other key components are all high-end special components to ensure high-precision, long-term stability, ultra-durable and low-maintenance equipment


RJ-45 network port, can be connected to the LIMS system


Low energy consumption, low noise, no need for external refrigerator, long-term stable and reliable operation


Security
Technical Specifications
Test MethodMeet SH/T 0770,ASTM D5972,IP 435
Test Range   
+20℃ ~ -80℃
Temperature resolution0.1℃
Temperature accuracy0.1℃
Repeatability0.5℃
Reproducibility0.8℃
Testing time~ 15min
Refrigeration methodPeltier/compressor hybrid refrigeration
Cooling rate15℃/min ± 5℃/min, adjustable
Heating rate10℃/min ± 0.5℃/min
Freezing point detectionOptical Sensors
Sample size0.14mL ~ 0.16mL
SecurityLogin password protection

Test end prompt

Use gas drying components, no compressed air required

User Interface12-inch color touch screen
Data storageNo less than 5,000
Communication interfaceUSB,RJ-45
Data OutputPrinter, LIMS
Machine power1000W
Instrument use environment

Temperature: 5~35℃;

Humidity: 20~80% RH

Instrument power supplyAC220V±10%,50Hz
Dimensions430mm wide × 510mm deep × 560mm high
Net weight of instrument37KG